Skip to content

U3 can be damaged by plugging in breakout board when power is on #9

@jonnew

Description

@jonnew

The PCIe controller with onix-fmc-host attached is inside a computer and powered on when the breakout board is plugged via the shrunk-delta ribbon cable, U3 can become degraded and eventually fail catastrophically. This never occurs if the breakout board is plugged when power is off.

image

  • Is there some way to keep this chip disabled until the breakout is detected?
  • Would this even help? Given that this is a reciever, we need top probe the input lines during a hot plug in event to see what they are doing.